3/Post Service Checkout
1503-0151-000
5/26/0
3-1
After servicing the ventilator, run any service menu tests that are pertinent to the
components replaced; Perform calibration on Flow sensors, Pressure sensitivity,
Flow valve and Bleed resistor. You must complete the checkout procedure for the
entire system; the ventilator, the anesthesia system, and all the accessories and
options.
WARNING:
w
w
w
w
You must perform all post-service checks after maintenance or service
of the ventilator. Failure to do so may result in patient injury.
WARNING:
w
w
w
w
All components and accessories must be connected correctly. All hoses
and cables must be properly connected before returning the anesthesia
system to clinical use. Failure to do so may result in patient injury.
3.1. Test the Ventilator.
Perform section "Preoperative checkout procedure" found in the Anesthesia Sys-
tem Operation and Maintenance Manual.
3.2. Test the Anesthesia System.
The ventilator is an integral part of a complete anesthesia system. To be certain
the ventilator is functioning correctly it is necessary to test the entire system.
Please refer to the anesthesia system manuals.
Important
:
If you have an Ohmeda Excel SE Anesthesia System, complete the
system checkout procedure in the Excel SE Service Manual. If you
have an Ohmeda Modulus® SE Anesthesia System, complete the
system checkout procedure in the Modulus® SE Service Manual.
3.3. Test all Options and Accessories.
Complete the checkout procedures for all options and accessories included with
the anesthesia system.
Summary of Contents for Aestiva 7900 SmartVent
Page 2: ...1503 0151 000 3 27 97 ...
Page 4: ...1503 0151 000 3 27 97 ...
Page 12: ...Notes viii 1503 0151 000 3 27 97 ...
Page 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Page 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Page 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...