6/ Repair Procedures
6-32
1503-0151-000
5/26/0
Removing the solenoid and gas inlet valve assembly
1. Solenoid socket head mounting screws 1.5 mm hex wrench
2. Cable connector to J6 microcontroller board
3. Captive mounting screws
4. Cable connector from pressure switch
Figure 6-27
Inlet valve and solenoid switch removal
1. Disconnect the cable connector (J6) from the solenoid switch on the microcon-
troller board.
Removing the GIV assembly
2. Use the 3 mm hex wrench to loosen the two (2) captive inlet valve mounting
screws.
3. Lift the inlet valve/solenoid switch assembly out of the ventilator chassis.
4. Examine the seat and O-rings in the manifold for damage. Replace as neces-
sary.
Removing the solenoid switch
Use a 1.5 mm hex wrench to remove the mounting screws from the solenoid switch
on the inlet valve.