6/ Repair Procedures
6-28
1503-0151-000
5/26/0
Power cord inlet removal
Figure 6-23
Power cord inlet connector removal
1. Item 1, figure 6-23 is the a-c inlet power connector where the two, 2 ampere
line fuses are located. Fuses may be replaced without removal of the inlet con-
nector.
2. Disconnect the white wire quick-connect terminal from the inlet connector as-
sembly to the transformer. (Rear connection, see figure 6-23 item 4.)
3. Disconnect the ground wire quick-connect terminal from the inlet connector as-
sembly to the ground lug on power module. (Middle connection.)
4. Disconnect the black wire quick-connect terminal from the inlet connector as-
sembly to the transformer. (front connection, see figure 6-23 item 3.)
5. Depress the locking tabs on both sides of the assembly and lift the assembly
out of the mounting hole in the power module.