2/Theory of Operation
2-2
1503-0151-000
3/27/97
Figure 2-2
7900 Ventilator Operational Block Diagram
Some 7900 Ventilator Features
•
No secondary regulator or exhalation solenoids
•
An exhalation valve that modulates flow in the pressure mode rath-
er than being just on and off
•
Pressure and volume mode selectable by the operator
•
All pneumatic components are mounted on a single manifold.
•
Each component is individually accessible from above or below the
manifold
•
There are no threaded connections to the subassemblies
•
Drive gas and bellows pressure relief valve gases are combined
and pass through the ventilator exhalation valve.
•
Exhalation valve block is autoclavable
•
Scavenging line runs from the ventilator to the Anesthesia Gas
Scavenging Receiver (AGSR) which scavenges both drive gas and
gas released by the bellows pressure relief valve.
•
Operates in a "Closed-Loop" configuration during both volume and
pressure modes of operation
•
Easier to service, fewer components and improved performance.
Mechanical
Overpressure
Bleed Off
(MOBO)
Pressure
Switch
Drive Gas Check Valve
(DGCV)
Free
Breathing
Valve
Supply
Gas
Hose
Manifold
Pressure
Transducer
Flow
Valve
Pressure
Regulator
Inlet
Filter
Gas
Inlet
Valve
(GIV)
35 - 100 psi
Supply
25 psi
25 mm port
17 mm port
Exhaust
to Scavenging
System
Exhaust to
Ambient
Bleed Exhaust
to Ambient
Exhalation
Manifold
Summary of Contents for Aestiva 7900 SmartVent
Page 2: ...1503 0151 000 3 27 97 ...
Page 4: ...1503 0151 000 3 27 97 ...
Page 12: ...Notes viii 1503 0151 000 3 27 97 ...
Page 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Page 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Page 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...