6/ Repair Procedures
6-30
1503-0151-000
5/26/0
Removing the flow control valve
1. Disconnect the cable connector (J7) from microcontroller board.
Figure 6-25
Disconnect the flow control valve cable
2. Use the 3 mm hex wrench to loosen the two (2) captive mounting screws and
lift out the flow control valve. (See figure 6-26)
3. Examine the seat and two O-rings for damage; replace as necessary.
Summary of Contents for Aestiva 7900 SmartVent
Page 2: ...1503 0151 000 3 27 97 ...
Page 4: ...1503 0151 000 3 27 97 ...
Page 12: ...Notes viii 1503 0151 000 3 27 97 ...
Page 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Page 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Page 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...