6/ Repair Procedures
6-18
1503-0151-000
5/26/0
To install 7900 software field upgrade
The software in the Cable Software Upgrade kit must be installed before follow-
ing these instructions.
1. Turn on the computer and enter Windows.
2. Insert the 7900 field upgrade installation diskette in drive A.
3. From the File menu of the Program Manager or File Manager, choose Run.
4. Type a:setup
5. Follow the Setup instructions on the screen.
6. Run the virus scanning software to confirm no known viruses exist on the computer.
To perform software upgrade
1. Turn off power to the 7900 ventilator.
2. Double-click on the Ohmeda Upgrade Tool icon located in the Ohmeda Sup-
port Tools program group.
3. From the Setup menu of the Ohmeda Upgrade Tool, choose release. Double
click on the file corresponding to the new release, for example, 7900_2_2.rls is
7900 software revision 2.2. Verify Comm Port is correct.
4. There are two choices, Step-by-Step or Quick Upgrade. Choose Step-by-Step
and follow the instructions on the screen. Confirm the serial port used by the
PC and the software version being loaded. If these are incorrect, choose Can-
cel and use the Setup menu to change these settings.
5. Continue to follow instructions on the screen until the upgrade procedure is com-
plete. Once you are familiar with the upgrade procedure, the Quick Upgrade but-
ton allows you to perform the upgrade without the step-by-step instructions. You
should follow the step-by-step process at least once for each software release,
since the procedures may change from one release to the next.
When you have finished with the upgrade, perform the tests listed below for replac-
ing the microcontroller board and/or EPROMs.
Microcontroller board
Whenever you replace the microcontroller board and/or the EPROM’s (U30 and
U40), or perform the software upgrade, perform the following tests in the ventilator
service mode.
Select the drive gas
Select the altitude
Cal O2 Sensor
Cal Flow Sensor
Cal Pressure Sensitivity
Cal Flow Valve
Cal Bleed Resistor
Perform the Post Service Checkout of the ventilator and Anesthesia system.
Summary of Contents for Aestiva 7900 SmartVent
Page 2: ...1503 0151 000 3 27 97 ...
Page 4: ...1503 0151 000 3 27 97 ...
Page 12: ...Notes viii 1503 0151 000 3 27 97 ...
Page 18: ...1 Introduction 1 6 1503 0151 000 5 26 0 Notes ...
Page 44: ...2 Theory of Operation 2 26 1503 0151 000 3 27 97 Notes ...
Page 46: ...3 Post Service Checkout 3 2 1503 0151 000 5 26 0 Notes ...