6/ Repair Procedures
6-14
1503-0151-000
5/26/0
Removing Microcontroller circuit board
CAUTION:
w
Disconnect the internal battery before attempting to remove or repair any
circuit board. Failure to do so may damage the internal electronics.
1. Pneumatic quick-connect fitting from main manifold
2. Connector from pressure sensor switch (J8)
3. Connector from GIV solenoid (J6)
4. connector from flow valve (J7)
Figure 6-12
Microcontroller board connector and cable identification around manifold pres-
sure transducer