CHAPTER 6 IMAGE FORMATION SYSTEM
6-82
COPYRIGHT © 2002 CANON INC. CANON CLC1100/1130/1150/1160/1180 REV.2 FEB. 2002 PRINTED IN JAPAN (IMPRIME AU JAPON)
3. Removing the Gradation Sensor
(CLC1140/1160/1180 only)
1) Disconnect the 2 connectors [1] and
remove the screw [2]; then, detach the
black toner concentration sensor unit [3]
from the machine.
Figure 6-676a
2) Remove the 2 screws [4].
3) Remove the 2 screws [5], and detach the
gradation sensor from the unit; then,
disconnect the connector [6].
Figure 6-676b
Figure 6-676c
•
When disconnecting the connector of the gradation sensor, be sure that the black
toner concentration sensor is ready to be detached from the unit to avoid an open
circuit.
• If you clean the gradation sensor, use a flannel cloth, moistened with and then
well wrung.
The reason for using water is to remove charges from the sensor window. (The
presence of charges can attract toner to the sensor window)
After cleaning the sensor window with water, do NOT leave droplets of water
on it, also do NOT dry wipe it
•
If you have replaced the gradation sensor, be sure to execute the following in service
mode:
FUNCTION>SNS-ADJ>P-SENS. Then, be sure also to record the readings of the
following in service mode on the Service Label:
ADJUST>DENS>P-SENS-Y/M/C/K.
[2]
[3]
[1]
[4]
[5]
[6]
Summary of Contents for CLC 1130
Page 8: ......
Page 24: ......
Page 26: ......
Page 33: ......
Page 55: ......
Page 101: ......
Page 103: ......
Page 121: ......
Page 141: ......
Page 143: ......
Page 159: ......
Page 239: ......
Page 247: ......
Page 255: ......
Page 261: ......
Page 263: ......
Page 369: ......
Page 397: ......
Page 399: ......
Page 451: ......
Page 453: ......
Page 493: ......
Page 495: ......
Page 549: ......
Page 556: ......
Page 563: ......
Page 586: ......
Page 670: ......
Page 674: ......
Page 683: ......
Page 698: ......
Page 704: ......
Page 711: ......
Page 785: ......
Page 799: ......
Page 829: ......
Page 945: ......
Page 957: ......
Page 959: ......
Page 963: ......