44
STAND-ALONE CONFIGURATION
4
POWERING ON SEQUENCE FOR THE VENTILATOR
Power on the ventilator by pressing the power
button for the ventilator.
Ensure the green power indicator light on the ventilator is on.
When the startup screen is displayed, listen for audible tones to test
the ventilator's alarm speaker.
During the ventilator’s start up sequence, the ventilator will perform a self
test. During the test, all ventilator indicator lights should briefly flash and an
audible alarm should briefly sound. This self test can take up to 15 seconds
to complete. If you do not hear tones when you turn on the ventilator, contact
your Breathe Technologies® service representative.
When the
Home
Screen is displayed, the touch screen is ready to use.
NOTE:
Ventilation will not begin until an Activity Button is selected on the
ventilator. For more information see "Choosing an Activity Button (Patient-
Selectable) to Begin Ventilation" on page 74.
1
2
CONNECTING AN INTERFACE TO THE VENTILATOR IN STAND-ALONE
CONFIGURATION
Plug the Breathe Pillows Interface™ or Breathe Universal Circuit
™
connector
into the interface port on the bottom of the ventilator until it clicks. For
more information about wearing interfaces, see "Chapter 5: Connecting an
Interface" on page 51.
NOTE:
Ensure that the Breathe Pillows Interface™ or Breathe Universal
Circuit
™
connector is connected to the ventilator in Extended Range
Configuration and Stand-Alone Configuration.