VENTILATION SETTINGS
6
SETTING BREATH TIMEOUT (APNEA BACKUP VENTILATION MODE)
On the
Clinician’s Setting
screen, touch
Breath Timeout
.
On the
Breath Timeout
screen, touch the
Timeout Period
or
Timeout Action
box you want to change.
TIP:
The
Breath Timeout
screen has two options for setting the
Breath Timeout
alarm. You can set the time to trigger the alarm at 20 or 60 seconds and the
backup ventilation mode to 3 LPM continuous flow or 12 BPM at the current
volume setting.
Touch the
Up
arrow or the
Down
arrow to change the value in the box.
Pressing and holding an arrow button will continue to increase or decrease the value.
Repeat steps 2 and 3 for each setting you want to change, and then press
OK
.
In the message asking if the settings are
OK
, touch
CONFIRM
.
NOTE:
Changes to settings only take effect when you touch
CONFIRM
.
BREATH TIMEOUT SETTINGS SUMMARY
PARAMETER
DEFAULT
ALTERNATE
Timeout Period
60 seconds
20 seconds
Timeout Action
12 BPM
3 LPM
10:11a, Fri
Jun 12, 2015
Clinician’s Settings
Ventilation
Settings
Alarm
Limits
Breath
Timeout
Source
Gas
1
10:11a, Fri
Jun 12, 2015
Breath Timeout
10:11a, Fri
Jun 12, 2015
20
Sec
12
BPM
2
3
3
10:11a, Fri
Jun 12, 2015
4
10:11a, Fri
Jun 12, 2015
Breath Timeout
5
10:11a, Fri
Jun 12, 2015
Values displayed on the screen are for
illustrative purposes only.