CHAPTER 8: MAINTENANCE
CLEANING BEFORE FIRST USE
It is not necessary to clean or sterilize the Life2000® Ventilation System before the first use.
DAILY CHECKS
Look at the ventilation system components daily. If any of the following conditions are discovered, discontinue use
of the ventilation system:
• Check for cracks in the casing.
• Check for loose or damaged buttons, connectors, or other control and alarm components.
• Check the interface and the source gas supply hose (if applicable) for leaks and loose or damaged cabling
or connectors.
Daily, or more often if necessary, check and empty the compressor’s condensation tray and replace if necessary.
For more information, see "Checking and Replacing the Condensation Tray" on page 115.
Essential Performance:
• Absence of system fault alarms
• Unintended change of settings and modes
• Absence of false alarms
• No interruption of operation without alarms.
If damage is discovered, discontinue use of the ventilation system or do not begin using the ventilation system.
For instructions on servicing or replacing damaged ventilation system components, contact your Breathe
Technologies® service representative.
ENVIRONMENTAL SPECIFICATIONS
Do not use the ventilation system if the ambient temperature is greater than 40°C (104°F) or less than 5°C (41°F).
Store the ventilation system in ambient temperatures less than 60°C (140°F) and greater than -20°C (-4°F).
WARNING:
The backside of the ventilator enclosure may reach 49°C in a 40°C environment.