MAINTENANCE
CLEANING FOR MULTI-PATIENT USE
WARNING:
To prevent risk of cross-contamination, clean and disinfect the ventilation system before using it on a new
patient, and use a new Breathe Pillows Interface™ or Breathe Universal Circuit
™
connector. For the third-party
patient mask, refer to the user guide provided by the manufacturer. Replace the oxygen hose between patients.
In addition to the cleaning and maintenance instructions for single-patient use, you must perform the following
before the ventilation system is provided to a new patient.
VENTILATION SYSTEM
Follow the instructions in "Cleaning and Disinfecting the Ventilation System" on page 109.
Replace the compressor’s condensation tray. For more information, see "Checking and Replacing the
Condensation Tray" on page 115.
Replace the compressor’s air inlet filter. For more information, see "Checking and Replacing the Air Inlet
Filter" on page 116.
Replace the compressor’s cooling filter assembly. For more information, see "Checking and Replacing the
Cooling Filter Assembly" on page 117.
OXYGEN HOSES
Replace between patients.
BREATHE PILLOWS INTERFACE™
Replace between patients.
BREATHE UNIVERSAL CIRCUIT™ CONNECTOR
Replace between patients.
THIRD-PARTY PATIENT MASK OR TUBE
Follow the manufacturer’s instructions.