71
6
VENTILATION SETTINGS
SETTING ALARM LIMITS FOR BREATH RATE AND PIP
To view or edit critical alarms, access the
Alarm Limits
screen from the
Clinician’s Settings
screen.
On the
Clinician’s Setting
screen, touch
Alarm Limits
.
On the
Set Alarm Limits
screen, touch the box corresponding to the
alarm limit you want to change.
Touch the
Up
arrow or the
Down
arrow to change the value in the box.
Pressing and holding an arrow button will continue to increase or decrease
the value.
Repeat steps 2 and 3 for each setting you want to change, and then press
OK
.
In the message asking if the settings are
OK
, touch
CONFIRM
.
NOTE:
Changes to settings only take effect when you touch
CONFIRM
.
ALARM LIMITS SETTINGS SUMMARY
ALARM
DEFAULT
MINIMUM
MAXIMUM
INCREMENT
High Breath Rate Alarm Limit (BPM) 50
5
120
1
Low Breath Rate Alarm Limit (BPM) 5
0
119
1
High PIP Alarm Limit (cmH
2
O)
20
5
40
1
Low PIP Alarm Limit (cmH
2
O)
1
1
15
1
TIP:
The breath rate monitor value is based on a four-breath average.
10:11a, Fri
Jun 12, 2015
Clinician’s Settings
Ventilation
Settings
Alarm
Limits
Breath
Timeout
Source
Gas
10:11a, Fri
Jun 12, 2015
1
10:11a, Fri
Jun 12, 2015
Alarm Limits
2
3
3
10:11a, Fri
Jun 12, 2015
4
Values displayed on the screen are for
illustrative purposes only.