20
STATIONARY CONFIGURATION
2
POWERING ON SEQUENCE IN STATIONARY CONFIGURATION
Power on the compressor by pressing the power button for the
compressor.
When powered on, lights surrounding the compressor power button
will illuminate:
Green indicates the compressor is connected to AC power.
Orange indicates the internal battery is being used.
NOTE:
If the ventilator is properly docked, the locked icon will also
illuminate when the compressor is powered on.
Power on the ventilator by pressing the power button for the ventilator.
Ensure the green power indicator light on the ventilator is on.
When the startup screen is displayed, listen for audible tones to test the
ventilator's alarm speaker.
During the ventilator’s start up sequence, the ventilator will perform a self
test. During the test, all ventilator indicator lights should briefly flash and an
audible alarm should briefly sound. This self test can take up to 15 seconds
to complete. If you do not hear tones when you turn on the ventilator,
contact your Breathe Technologies® service representative.
When the
Home
Screen is displayed, the touch screen is ready to use.
NOTE:
Ventilation will not begin until an Activity Button is selected on the
ventilator. For more information see "Choosing an Activity Button (Patient-
Selectable) to Begin Ventilation" on page 74.
NOTE:
A high-pitched sound may be heard briefly while the compressor is
powering on. This sound will go away once the compressor reaches operating
pressure.
0
100%
1
2
0
100%
4
3