74
VENTILATION SETTINGS
6
CHOOSING AN ACTIVITY BUTTON (PATIENT-SELECTABLE) TO BEGIN VENTILATION
When the ventilator is first powered on, you must select an Activity Button to begin ventilation. The three Activity
Buttons on the ventilator are programmed to correspond to up to three different prescriptions as directed by a
physician.
High Activity Button
Medium Activity Button
Low Activity Button
Choose an Activity Button appropriate for the patient’s needs. The selection can be changed by the patient at any
time, if set and activated by a clinician.
NOTE:
One, two, or three Activity Buttons may be available, as directed by a physician. For more information, see
"Prescription Settings" on page 60.
Ensure that the ventilator is powered on.
Ensure a pressure source (the compressor, an oxygen cylinder, or an oxygen wall source) is connected to
the ventilator and turned on.
Press and hold an Activity Button until you hear a tone that
indicates it is active. The touch screen will display the
Home
Screen
and the ventilator will begin delivering therapy.
Confirm the selected
Activity Icon
is displayed at the bottom of the touch
screen and the icon and volume are displayed at the top of the screen.
(The
High
Activity Icon
is shown for illustrative purposes here.) The
ventilator will begin ventilating using the chosen prescription parameters for
the next breath.
NOTE:
The currently-ventilating
Activity Icon
and volume are displayed at
the top of all screens unless you are in the Clinician’s Menu.
3
20.0
7.5
15
500 ml
Air LPM
PIP cmH
2
O
Breath/min
10:11a, Fri
Jun 12, 2015
4
4
Values displayed on the screen are for
illustrative purposes only.