141
11
PERFORMANCE SPECIFICATIONS
PARAMETER
DESCRIPTION
Storage and Transport Environment
Storage & Transport
Temperature
-20°C to +60°C (-4°F to +140°F)
Storage & Transport
Humidity
10% to 95% non-condensing
Storage & Transport
Altitude
1100 hPa to 624 hPa or -2290 ft. to +8000 ft. (-700 m to +2439 m)
Operating Environment
Operating Temperature
+5°C to +40°C (+41°F to +104°F)
Operating Humidity
10% to 95% non-condensing
Operating Altitude
1100 hPa to 624 hPa or -2290 ft to +8000 ft (-700 m to +2439 m)
NOTES:
• The performance of the compressor may degrade in high temperature, high humidity, or high altitude
environments. If degradation is seen, switch to an alternate means of ventilation.
• The ventilator settings might not be achieved
when sourced by the Life2000
®
Compressor
due to increases in altitude near or above 2500
feet. Consult the table to ensure that the
compressor can meet the ventilator settings.
The tidal volume delivered to the patient
includes the ventilator set entrainment
volume from patient int supplemental
oxygen volume (if used). For additional
information see page 152.
Physical Features
Ventilation system weight
16 lb.
Ventilation system size
12.1” x 8.7” x 8.6”
Ventilator weight
1.1 lb.
Ventilator size
3.2” x 7.7” x 1.0”
Compressor weight
14.9 lb.
Compressor size
12.1” x 8.7” x 8.6”
Latex
The ventilation system does not contain natural rubber latex.
Expected Service Life
Expected Service Life
The Life2000® Ventilation System has a five year design life when operated in
accordance with the
Instructions For Use
provided.
Do not use the ventilation system past its expected service date.
The battery shall be replaced if performance degrades to an unacceptable level
during the one year warranty period.
Service Intervals
Service Intervals
1 year and 2.5 years from the date of shipment
Simulated Elevation
(in feet)
Observed Maximum
Compressor Output (in LPM)
0
17
2500
14
4000
12
8000
8