PERFORMANCE SPECIFICATIONS
Breathe Universal Circuit
™
connector
Used to connect commercially available non-invasive masks (full face, nasal, and
pillows) or tracheostomy tubes for invasive ventilation
Acoustic levels during normal operation
Compressor
≤ 60 dBA at a distance of 1 meter
PARAMETER
DESCRIPTION
Ventilator Alarm Notifications
High Priority Alarm
At Maximum loudness setting: 73 db(A)
At Minimum loudness setting: 60 db(A)
Medium Priority Alarm
At Maximum loudness setting: 72 db(A)
At Minimum loudness setting: 59 db(A)
Ventilator Alarms
High Circuit Pressure
Interface may be pinched or kinked.
High PEEP Pressure
Interface may be blocked.
High PIP Pressure
Peak Inspiratory Pressure (PIP) exceeds the set limit.
High Temperature
Ventilator CPU or battery temperature is above the allowable limit.
Very Low Battery
Ventilator battery capacity drops below 15%.
Battery Low
Ventilator battery capacity drops below 25%.
Breath Timeout
No breath is detected for 20 seconds or 60 seconds, depending on the setting.
High Breath Rate
Respiratory rate exceeds the set limit.
High Delivery Pressure
Interface pressure during delivery exceeds the maximum expected.
High Gas Pressure
Source gas pressure exceeds the allowable limit (95 PSI).
Low Breath Rate
Respiratory rate falls below the set limit.
Low Delivery Pressure
Interface pressure during delivery fails to exceed the minimum expected.
Low Gas Pressure
Source gas pressure drops below the allowable limit (35 PSI).
Low PIP Pressure
PIP below set limit
System Fault
Internal fault detected during operation.
POST System Fault
System fault detected during ventilator power on.
Monitors and Indicators
Ventilator user interface
Touch screen LCD for settings, monitors, and alarms.
Mechanical buttons for on/off, alarm silence, and prescription settings
Pressure monitoring
Ventilator output pressure and airway pressure are monitored continuously. Airway
pressure is monitored using a lumen in the gas delivery circuit wall, which terminates
in a sensing port on the distal end of the circuit.
Volume Setting displays
Delivered volume setting is displayed.
Ventilator breath indicator
An LED visual indicator indicating an inspiration.
Compressor battery
indicator
When the compressor is powered on, the internal battery level is displayed using
indicator lights in the battery charge scale. An alarm occurs when battery capacity is
20% or less.