18
STATIONARY CONFIGURATION
2
DOCKING THE VENTILATOR INTO THE COMPRESSOR
NOTE:
When switching between ventilation system configurations, ensure that the patient receives adequate
ventilation therapy; place the patient on an alternate means of ventilation if necessary.
Ensure that the ventilator is powered off.
NOTE:
Alarms might be encountered and/or the selected Activity
Button might be inadvertently changed if the ventilator is not
powered off before docking.
To dock the ventilator into the compressor, make sure the locking
knob is in the unlock position as shown.
Position the ventilator into the compressor cradle as shown. Push the
ventilator into the compressor in the direction indicated by the arrow
until it clicks into place.
Gently push the top of the ventilator (in the area indicated) until the
front of the ventilator is flush with the front of the compressor and
clicks into place.
TIP:
Avoid pushing directly on the touch screen or buttons.
1
0
100%
2
Push here