69
6
VENTILATION SETTINGS
To set a
Prescription Setting
for Assist ventilation mode, the following parameters need to be set according to the
table below:
VENTILATION MODE*
SET TRIGGER SENSITIVITY TO:
SET BREATH RATE TO:
Assist
0 to 9
0
* SETTINGS BASED ON PRESCRIPTION
To set the
Prescription Setting
:
On the
Clinician’s Setting
screen, touch
Ventilation Settings
.
On the
Ventilation Settings
screen, touch the box beside the setting you
want to change.
Touch the
Up
arrow to increase the value in the box or the
Down
arrow to
decrease it. Pressing and holding an arrow button will continue to increase or
decrease the value.
Repeat steps 2 and 3 for each setting you want to change, and then press
OK
.
In the message asking if the settings are
OK
, touch
CONFIRM
.
NOTE:
Changes to settings only take effect when you touch
CONFIRM
.
10:11a, Fri
Jun 12, 2015
Clinician’s Settings
Ventilation
Settings
Alarm
Limits
Breath
Timeout
Source
Gas
1
10:11a, Fri
Jun 12, 2015
Ventilation Settings
10:11a, Fri
Jun 12, 2015
ml
FLIP
CANCEL
OK
Volume
500
I-Time
PEEP
BR
sec
cmH
2
O
/min
1.00
4
0
0-9, OFF
Sensitivity
4
3
3
2
4
10:11a, Fri
Jun 12, 2015
10:11a, Fri
Jun 12, 2015
1
FLIP
OK
CANCEL
Are settings OK?
CONFIRM
BACK
Ventilation Settings
5
Values displayed on the screen are for
illustrative purposes only.