104
ALARMS, ALERTS, AND TROUBLESHOOTING
7
ADDITIONAL TROUBLESHOOTING SITUATIONS (CONTINUED)
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Breath indicator light is
not flashing and there is
no volume output
Ventilator is not on.
Ensure ventilator is on.
Ensure ventilator is on.
Ensure ventilator is on.
A ventilator Activity Button has not
been pressed.
Press an Activity Button on the ventilator.
Press an Activity Button on the ventilator.
Press an Activity Button on the ventilator.
Ventilator is delivering
gas without being
triggered by patient
effort.
Ventilator is in Control or
Assist/Control ventilation mode.
It is normal for the ventilator to deliver therapy
based on breath rate and breath timeout settings;
adjust settings as required.
It is normal for the ventilator to deliver therapy
based on breath rate and breath timeout settings;
adjust settings as required.
It is normal for the ventilator to deliver therapy based on
breath rate and breath timeout settings; adjust settings as
required.
The trigger sensitivity is too
sensitive.
Adjust trigger sensitivity to a higher number (lower
sensitivity).
Adjust trigger sensitivity to a higher number (lower
sensitivity).
Adjust trigger sensitivity to a higher number (lower sensitivity).
Secretions have built up on the
interface.
Inspect and clean the interface per the instructions
for cleaning the interface.
Inspect and clean the interface per the instructions
for cleaning the interface.
Inspect and clean the interface per the instructions for
cleaning the interface.
Patient interface or connector is
damaged.
Check the patient interface and connector for
damage and replace if necessary.
Check the patient interface and connector for
damage and replace if necessary.
Check the patient interface and connector for damage and
replace if necessary.
Ventilator sometimes
misses breaths.
Patient is breathing faster than 40
BPM.
It is normal for the ventilator to limit breath rate to
less than 40 BPM.
It is normal for the ventilator to limit breath rate to
less than 40 BPM.
It is normal for the ventilator to limit breath rate to less than 40
BPM.
Secretions have built up on the
interface.
Inspect and clean the interface per the instructions
for cleaning the interface.
Inspect and clean the interface per the instructions
for cleaning the interface.
Inspect and clean the interface per the instructions for
cleaning the interface.
Patient interface is not connected or
is leaking.
Verify that the interface is properly connected to
the ventilator and is not leaking at patient side.
Verify that the interface is properly connected
to the ventilator (not the compressor) and is not
leaking at patient side.
Verify that the interface is properly connected to the
compressor and is not leaking at patient side.
Ventilator is triggering
during exhalation.
Secretions have built up on the
interface.
Inspect and clean the interface per the instructions
for cleaning the interface.
Inspect and clean the interface per the instructions
for cleaning the interface.
Inspect and clean the interface per the instructions for
cleaning the interface.
Therapy delivery is
causing coughing or
irritation in airway.
Interface is not positioned correctly. Reposition the interface per the instructions in
"Chapter 5: Connecting an Interface".
Reposition the interface per the instructions in
"Chapter 5: Connecting an Interface".
Reposition the interface per the instructions in "Chapter 5:
Connecting an Interface".
Patient is breathing against
ventilator-triggered breaths.
Switch to another Activity Button.
If symptoms persist, contact your physician.
Switch to another Activity Button.
If symptoms persist, contact your physician.
Switch to another Activity Button.
If symptoms persist, contact your physician.
Ventilator battery does
not last as long as
expected after a charge.
Ventilator battery is not charged
completely.
Connect the ventilator to the ventilator battery
charger and an AC power source or dock the
ventilator back into the compressor and power on
the compressor to recharge the battery.
Connect the ventilator to the ventilator battery
charger and an AC power source or dock the
ventilator back into the compressor and power on
the compressor to recharge the battery.
Recharge the ventilator battery. Check that the ventilator is
correctly docked in the compressor and make sure that the
ventilator charging indicator light (the locked icon) on the
compressor is illuminated when the compressor is on.
Ventilator battery life is nearing its
end.
Contact your Breathe Technologies® service
representative.
Contact your Breathe Technologies® service
representative.
Contact your Breathe Technologies® service representative.
Ventilator screen is
unresponsive
Alarm Silence button
may be stuck.
Contact your Breathe Technologies® service
representative.
Contact your Breathe Technologies® service
representative.
Contact your Breathe Technologies® service
representative.