STATIONARY CONFIGURATION
2
SILENCE ALARM BUTTON ON THE COMPRESSOR
Silencing and clearing alarms is a multi-step process that depends on alarm priority and how many alarms are
active. For more information see "Chapter 7: Alarms, Alerts, and Troubleshooting".
Press the Silence Alarm button to temporarily
silence the alarm for 60 seconds. Pressing the
Silence Alarm button silences only one alarm at a
time—in audible or vibrating alarm mode. If more
than one alarm occurs, press the Silence Alarm
button once for each alarm. If the alarm is a
medium- or high-priority alarm and is not silenced
after 60 seconds, the alarm will continue with an
additional buzzer.
Resolve the condition that triggered the alarm. For help resolving alarms, see the alarm and troubleshooting
tables in "Chapter 7: Alarms, Alerts, and Troubleshooting" for possible causes of an alarm and options to
resolve it. If an alarm silence button is pressed but the condition that triggered the alarm is not resolved, the
alarm will sound again after 60 seconds.