98
ALARMS, ALERTS, AND TROUBLESHOOTING
7
LOW-PRIORITY ALARMS
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
POST System Fault
10:11a, Fri
Jun 12, 2015
--
Breath/min
LPM
PIP cmH
2
O
--
--
POST System Fault
System fault is detected during
ventilator power on.
Power off the ventilator, and power it on again.
If the system fault persists, place the patient on
an alternate means of ventilation (if necessary)
and contact your Breathe Technologies® service
representative.
Power off the ventilator, and power it on again.
If the system fault persists, place the patient on
an alternate means of ventilation (if necessary)
and contact your Breathe Technologies® service
representative.
Power off the ventilator, and power it on again.
If the system fault persists, place the patient on an alternate
means of ventilation (if necessary) and contact your Breathe
Technologies® service representative.
MEDIUM-PRIORITY ALARMS (CONTINUED)
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Low PIP Pressure
10:11a, Fri
Jun 12, 2015
13
Breath/min
Air LPM
PIP cmH
2
O
0.0
4.3
Low PIP Pressure
Peak Inspiratory Pressure (PIP)
below set limit
Ensure patient interface is not leaking at patient
side.
Switch to another active Activity Button.
If the alarm persists, contact your physician.
Ensure patient interface is not leaking at patient
side.
Switch to another active Activity Button.
If the alarm persists, contact your physician.
Ensure patient interface is not leaking at patient side.
Ensure the ventilator is properly docked into the
compressor.
Switch to another active Activity Button.
If the alarm persists, contact your physician.
System Fault
10:11a, Fri
Jun 12, 2015
--
Breath/min
O
2
LPM
PIP cmH
2
O
--
--
System Fault.
Press OK to
restart system.
System Fault
Internal fault detected during
operation
If a system fault occurs, in the message on the
touch screen to reboot, touch
OK
; the ventilator
will turn itself off and then on again. Restart
ventilation by pressing an Activity Button on the
ventilator.
If the system fault persists, place the patient on
an alternate means of ventilation (if necessary)
and contact your Breathe Technologies® service
representative.
If a system fault occurs, in the message on the
touch screen to reboot, touch
OK
; the ventilator
will turn itself off and then on again. Restart
ventilation by pressing an Activity Button on the
ventilator.
If the system fault persists, place the patient on
an alternate means of ventilation (if necessary)
and contact your Breathe Technologies® service
representative.
If a system fault occurs, in the message on the touch screen
to reboot, touch
OK
; the ventilator will turn itself off and then
on again. Restart ventilation by pressing an Activity Button
on the ventilator.
If the system fault persists, place the patient on an alternate
means of ventilation (if necessary) and contact your Breathe
Technologies® service representative.