62
VENTILATION SETTINGS
6
FACTORY DEFAULT PRESCRIPTION SETTINGS
The following table lists the Life2000® factory default
Prescription Settings
that may be edited. For a full list of
factory defaults, see the "Summary of Factory Default Settings" on page 84.
DESCRIPTION
DEFAULT VALUE
LOW
ACTIVITY
MEDIUM
ACTIVITY
HIGH
ACTIVITY
Ventilation Settings
Volume
100 ml
150 ml
190 ml
I-Time (Inspiratory Time)
0.75 sec
PEEP (Positive End Expiratory Pressure)
0 cmH
2
O
Sensitivity (Trigger Sensitivity)
4*
BR (Breath Rate)
12 BPM*
* Ventilation mode: Assist/Control ventilation mode
Alarm Limits
High BR (Breath Rate) alarm limit
50 BPM
Low BR (Breath Rate) alarm limit
5 BPM
High PIP (Peak Inspiratory Pressure) alarm limit
20 cmH
2
O
Low PIP (Peak Inspiratory Pressure) alarm limit
1 cmH
2
O
Breath Timeout
Breath Timeout Period
60 seconds
Breath Timeout Action
12 BPM
Source Gas
Source Gas
Air
NOTE:
Each
Prescription Setting
is independent of other
Prescription Settings
.
The following factory default cannot be edited:
DESCRIPTION
DEFAULT VALUE
High PEEP (Positive End Expiratory Pressure)
Pressure alarm limit
+7 cmH
2
O (above PEEP setting)