41
4
STAND-ALONE CONFIGURATION
UNDOCKING THE VENTILATOR FROM THE COMPRESSOR
NOTE:
When switching between ventilation system configurations, ensure that the patient receives adequate
ventilation therapy; place the patient on an alternate means of ventilation if necessary.
Power off the ventilator.
NOTE:
Alarms might be encountered and/or the selected Activity
Button might be inadvertently changed if the ventilator is not
powered off before undocking.
Turn the locking knob counterclockwise to the unlocked position until
the compressor ejects the ventilator.
NOTE:
Some tension will be encountered while turning during
ejection.
Pull the ventilator (in the direction shown by the arrow) to separate it
from the compressor cradle.
Power off the compressor.
NOTE:
For docking instructions, see "Docking the Ventilator into the Compressor" on page 18.
0
100%
1
2
0
100%
4
TESTING THE VENTILATOR
In a multi-patient setting, the ventilator must be tested before it is assigned to a new patient. For instructions on
testing the ventilator, see "Testing Ventilator Alarms" on page 118.