ALARMS, ALERTS, AND TROUBLESHOOTING
ADDITIONAL TROUBLESHOOTING SITUATIONS (CONTINUED)
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
An on-screen alarm
flashes on the screen
intermittently
Low gas pressure
The cylinder is nearly empty; connect the ventilator
to a full cylinder. If the issue persists, contact your
Breathe Technologies® service representative.
Contact your Breathe Technologies® service
representative.
Contact your Breathe Technologies® service
representative.
Ventilator buzzer
sounds constantly for
two to five minutes and
screen goes black
The ventilator battery is damaged.
Contact your Breathe Technologies® service
representative.
Contact your Breathe Technologies® service
representative.
Contact your Breathe Technologies® service representative.
Ventilator does not turn
on
Ventilator is not correctly docked in
the compressor.
N/A
N/A
Check that the ventilator is correctly docked in the
compressor and make sure that the ventilator charging
indicator light (the locked icon) on the compressor is
illuminated when the compressor is on.
Ventilator battery is completely
discharged.
Connect the ventilator to the ventilator battery
charger and an AC power source or dock the
ventilator back into the compressor and power
on the compressor to recharge the battery.
If the issue is not resolved, contact your Breathe
Technologies® service representative.
Connect the ventilator to the ventilator battery
charger and an AC power source or dock the
ventilator back into the compressor and power
on the compressor to recharge the battery.
If the issue is not resolved, contact your Breathe
Technologies® service representative.
Connect the compressor to an AC power source.
If the issue is not resolved, contact your Breathe
Technologies® service representative.
The source gas supply
hose does not connect
to the gas source.
An incorrect source gas supply
hose is being used.
Ensure that a Breathe Technologies
®
oxygen
hose is connected.
Ensure you are using a Breathe Technologies®
oxygen hose to connect the ventilator to the
compressor.
N/A
An incompatible regulator is being
used.
Ensure the gas regulator is 42-87 PSI with a
standard DISS fitting.
For any issues with the regulator, contact the
regulator manufacturer.
N/A
N/A
Cylinder does not last
as long as expected
User breath rate is higher than
expected.
Refer to "Cylinder Duration Information" on page
150.
N/A
N/A
Selected Activity Button requires
higher volumes of gas.
Obtain a new or larger cylinder.
N/A
N/A
Cylinder was not full at the
beginning of ventilation.
Obtain a new cylinder.
N/A
N/A
The gas regulator is not properly
connected to the gas cylinder.
Reconnect the gas regulator to the gas cylinder
and verify there are no leaks.
N/A
N/A
Flow regulator is on.
Ensure the flow valve is off or set to 0.
N/A
N/A
The gas regulator may have a leak. The regulator sealing washer that connects to the
gas cylinder may be worn or damaged. For any
issues with the regulator, contact the regulator
manufacturer.
N/A
N/A