STAND-ALONE CONFIGURATION
VENTILATOR SILENCE ALARM BUTTON
Silencing and clearing alarms is a multi-step process that depends on alarm priority and how many alarms are
active. For more information see "Chapter 7: Alarms, Alerts, and Troubleshooting".
Silence Alarm button on ventilator.
Press the Silence Alarm button to temporarily silence the alarm for 60
seconds. Pressing the Silence Alarm button silences only one alarm
at a time—in audible or vibrating alarm mode. If more than one alarm
occurs, press the Silence Alarm button once for each alarm. If the
alarm is a medium- or high-priority alarm and is not silenced after 60
seconds, the alarm will continue with an additional buzzer.
Resolve the condition that triggered the alarm. For help resolving
alarms, see the alarm and troubleshooting tables in "Chapter 7: Alarms, Alerts, and Troubleshooting" for
possible causes of an alarm and options to resolve it. If a Silence Alarm button is pressed but the condition
that triggered the alarm is not resolved, the alarm will sound again after 60 seconds.
After resolving a High Temperature, High Circuit Pressure, or High PEEP
Pressure high-priority alarm, touch
OK
in the message that indicates the
alarm has been resolved.
1
500 ml
High Circuit
Peak Pressure
has been resolved.
10:11a, Fri
Jun 12, 2015
3
Values displayed on the screen are for
illustrative purposes only.