ALARMS, ALERTS, AND TROUBLESHOOTING
7
MEDIUM-PRIORITY ALARMS
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Battery Low
10:11a, Fri
Jun 12, 2015
12
Breath/min
Air LPM
PIP cmH
2
O
3.9
1.1
Battery Low
Ventilator battery capacity drops
below 25%.
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge
the battery.
OR
Dock the ventilator back into the compressor to
recharge the battery. Ensure that the locked icon
on the compressor illuminates to indicate the
ventilator is charging. Ensure the compressor is
powered on.
If the battery does not recharge, contact your
Breathe Technologies® service representative.
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge
the battery.
OR
Dock the ventilator back into the compressor to
recharge the battery. Ensure that the locked icon
on the compressor illuminates to indicate the
ventilator is charging. Ensure the compressor is
powered on.
If the battery does not recharge, contact your
Breathe Technologies® service representative.
Ensure the ventilator is properly docked and locked into the
compressor and the locked icon is illuminated to indicate
the ventilator is charging.
Ensure the compressor is powered on.
If the battery does not recharge, contact your Breathe
Technologies® service representative.
Breath Timeout
10:11a, Fri
Jun 12, 2015
--
Breath/min
Air LPM
PIP cmH
2
O
1.4
3.0
Breath Timeout
No breath is detected for 20 or 60
seconds, depending on the setting
Patient is not breathing.
Patient is breathing through the mouth while
using the Breathe Pillows Interface™.
Breaths are too shallow to trigger ventilation;
change the trigger sensitivity to a lower setting
(higher sensitivity).
Ensure patient interface is not leaking at patient
side.
Inspect and clean the interface per the
instructions for cleaning the interface.
Patient is not breathing.
Check that the interface is connected to the
ventilator, not the compressor.
Patient is breathing through the mouth while
using the Breathe Pillows Interface™.
Breaths are too shallow to trigger ventilation;
change the trigger sensitivity to a lower setting
(higher sensitivity).
Ensure patient interface is not leaking at patient
side.
Inspect and clean the interface per the
instructions for cleaning the interface.
Patient is not breathing.
Patient is breathing through the mouth while using the
Breathe Pillows Interface™.
Breaths are too shallow to trigger ventilation; change the
trigger sensitivity to a lower setting (higher sensitivity).
Ensure patient interface is not leaking at patient side.
Inspect and clean the interface per the instructions for
cleaning the interface.
Ensure the ventilator is properly docked into the
compressor.
High Breath Rate
10:11a, Fri
Jun 12, 2015
43
Breath/min
Air LPM
PIP cmH
2
O
1.5
11.3
High Breath Rate
Respiratory rate exceeds the set
limit.
Patient is breathing faster than the rate set by the
clinician.
Inspect and clean the interface per the
instructions for cleaning the interface.
The ventilator may be false triggering because
the trigger sensitivity is set too low (in Assist
or Assist/Control ventilation mode). Verify the
ventilator is syncing with patient effort and adjust
trigger sensitivity higher (lower sensitivity).
Patient is breathing faster than the rate set by the
clinician.
Inspect and clean the interface per the
instructions for cleaning the interface.
The ventilator may be false triggering because
the trigger sensitivity is set too low (in Assist
or Assist/Control ventilation mode). Verify the
ventilator is syncing with patient effort and adjust
trigger sensitivity higher (lower sensitivity).
Patient is breathing faster than the rate set by the clinician.
Inspect and clean the interface per the instructions for
cleaning the interface.
The ventilator may be false triggering because the
trigger sensitivity is set too low (in Assist or Assist/Control
ventilation mode). Verify the ventilator is syncing with patient
effort and adjust trigger sensitivity higher (lower sensitivity).