140
PERFORMANCE SPECIFICATIONS
11
Ventilator battery indicator
Indicates percentage of battery charge remaining. Always displayed on touch
screen. Alarm occurs when capacity is below 25% and again at 15%.
Standards and Regulatory Compliance
AAMI/ANSI 60601-1:2005, ANSI/AAMI HE75:2009, AIM 7351731, ASTM F1246-91 (2005), IEC 60601-1-2:2007,
IEC 62133:2012 (2nd edition), IEC 62366:2007, ISO 10993-1:2009, ISO 80601-2-12:2011, ISO 80601-2-72:2015
PARAMETER
DESCRIPTION
Ventilator AC Battery Charger Specifications
Input AC voltage
100-240 VAC
Input AC frequency
50-60 Hz
Input AC current
0.3 A max.
Output DC voltage
8.4 VDC
Output DC current
1.3 A
Insulation class
Class II
Electrical safety approvals
UL 60601-1, EN 60950, EN 60601-1, EN 60335-2-29
Dimensions
3.55” x 1.77” x 1.26” (90 x 45 x 32 mm)
Weight
0.25 lb. (115 g)
Ventilator Power Cord Specifications
Cord length
6 ft.
Compliance
IEC 320-E7
Compressor Power Supply Specifications
Input AC voltage
100-240 VAC
Input AC frequency
50-60 Hz
Input AC current
5 A max. at 100 VAC
2.5 A max. at 240 VAC
Output DC voltage
24 VDC
Output DC current
12.5 A
Insulation class
Class II
Ingress protection rating
IP22
Electrical safety approvals
IEC 60601-1, IEC 60601-1-11
Dimensions
7.8” x 4” x 2” (198 x 102 x 51 mm)
Weight
3 lb. (1.36 kg)
Compressor Power Cord Specifications
Cord length
6 ft.
Compliance
IEC 320-C13
Compressor Output
Pressure
58 PSI ±15% (at sea level and normal ambient conditions)
Flow
17
–
21 LPM (at sea level and normal ambient conditions)
Outlet Fitting
9/16" - 18 male threaded fitting