94
ALARMS, ALERTS, AND TROUBLESHOOTING
7
MEDIUM-PRIORITY ALARMS (CONTINUED)
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
High Del. Pressure
(High Delivery
Pressure)
10:11a, Fri
Jun 12, 2015
14
Breath/min
Air LPM
PIP cmH
2
O
1.4
4.0
High Del. Pressure
Interface pressure during delivery
exceeds the maximum expected.
Check the interface; replace it if the tubing is
torn, bent, or kinked.
Ensure the interface tubing is not pinched,
crushed, bent, or kinked.
Check the interface; replace it if the tubing is
torn, bent, or kinked.
Ensure the interface tubing is not pinched,
crushed, bent, or kinked.
Check the interface; replace it if the tubing is torn, bent, or
kinked.
Ensure the interface tubing is not pinched, crushed, bent, or
kinked.
Ensure the ventilator is properly docked into the
compressor.
High Gas Pressure
10:11a, Fri
Jun 12, 2015
12
Breath/min
Air LPM
PIP cmH
2
O
1.4
3.1
High Gas Pressure
Gas pressure exceeds the
allowable limit.
Place the patient on an alternate means of
ventilation (if necessary).
Ensure that you are using a 50-PSI (nominal)
regulator.
If the alarm is not resolved, power off
the ventilator and contact your Breathe
Technologies® service representative.
Discontinue use of the ventilator. Place the
patient on an alternate means of ventilation
(if necessary), or connect the ventilator to an
alternate pressure source (oxygen cylinder or
wall source) in Stand-Alone Configuration.
Power off the compressor and contact your
Breathe Technologies® service representative.
Discontinue use of the ventilator. Place the patient on an
alternate means of ventilation (if necessary), or undock the
ventilator from the compressor and connect to an alternate
pressure source (oxygen cylinder or wall source) in Stand-
Alone Configuration.
Power off the compressor and contact your Breathe
Technologies® service representative.
Low Breath Rate
10:11a, Fri
Jun 12, 2015
13
Breath/min
Air LPM
PIP cmH
2
O
1.5
4.1
Low Breath Rate
Respiratory rate falls below set
limit.
Patient is breathing through the mouth while
using the Breathe Pillows Interface™.
Breaths are too shallow to trigger ventilation;
change the trigger sensitivity to a lower setting
(higher sensitivity).
Ensure patient interface is not leaking at patient
side.
Inspect and clean the interface per the
instructions for cleaning the interface.
Check that the interface is connected to the
ventilator, not the compressor.
Patient is breathing through the mouth while
using the Breathe Pillows Interface™.
Breaths are too shallow to trigger ventilation;
change the trigger sensitivity to a lower setting
(higher sensitivity).
Ensure patient interface is not leaking at patient
side.
Inspect and clean the interface per the
instructions for cleaning the interface.
Ensure that the ventilator is properly docked into the
compressor.
The patient is breathing through the mouth while using the
Breathe Pillows Interface™.
Breaths are too shallow to trigger ventilation; change the
trigger sensitivity to a lower setting (higher sensitivity).
Ensure patient interface is not leaking at patient side.
Inspect and clean the interface per the instructions for
cleaning the interface.
Low Del. Pressure
(Low Delivery
Pressure)
10:11a, Fri
Jun 12, 2015
11
Breath/min
Air LPM
PIP cmH
2
O
0.0
1.2
Low Del. Pressure
Interface pressure during delivery
fails to exceed the minimum
expected.
Check the interface connections.
Check the interface; replace it if it is leaking.
Check that the interface is connected to the
ventilator, not the compressor.
Check the interface connections.
Check the interface; replace it if it is leaking.
Ensure the ventilator is properly docked into the
compressor.
Check the interface connections.
Check the interface; replace it if it is leaking.
Values displayed on the screens are for illustrative purposes only.