91
7
ALARMS, ALERTS, AND TROUBLESHOOTING
HIGH-PRIORITY ALARMS (CONTINUED)
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Very Low Battery
10:11a, Fri
Jun 12, 2015
12
Breath/min
Air LPM
PIP cmH
2
O
3.9
1.1
Very Low Battery
Ventilator battery capacity drops
below 15%
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge
the battery.
OR
Dock the ventilator back into the compressor to
recharge the battery. Ensure that the locked icon
on the compressor illuminates to indicate the
ventilator is charging. Ensure the compressor is
powered on.
If the battery does not recharge, place the
patient on an alternate means of ventilation
(if necessary) and contact your Breathe
Technologies® service representative.
Connect the ventilator to the ventilator battery
charger and an AC power source to recharge
the battery.
OR
Dock the ventilator back into the compressor to
recharge the battery. Ensure that the locked icon
on the compressor illuminates to indicate the
ventilator is charging. Ensure the compressor is
powered on.
If the battery does not recharge, place the
patient on an alternate means of ventilation
(if necessary) and contact your Breathe
Technologies® service representative.
Ensure the ventilator is properly docked and locked into the
compressor and the locked icon is illuminated to indicate
the ventilator is charging.
Ensure the compressor is powered on.
If the battery does not recharge, place the patient on an
alternate means of ventilation (if necessary) and contact your
Breathe Technologies® service representative.