VENTILATION SETTINGS
SELECTING THE SOURCE GAS
The Breathe Technologies® Life2000® Ventilation System uses
Air
as the factory set default source gas. If using
an oxygen cylinder, select
O
2
as the
Source Gas.
Source gas is oxygen
Stand-Alone
Configuration
with an oxygen cylinder
(or a wall oxygen pressure source)
Source gas is air
Stationary
Configuration
Extended Range Configuration
O
2
On the
Clinician’s Setting
screen, touch
Source Gas
.
On the
Source Gas
screen, touch the appropriate check box for the source
gas prescribed by the physician.
Touch
OK
to confirm your selection.
In the message asking if the settings are
OK
, touch
CONFIRM
.
NOTE:
Changes to settings only take effect when you touch
CONFIRM
.
SOURCE GAS SETTINGS SUMMARY
DEFAULT
ALTERNATE
Air
O
2
(Oxygen)
Clinician’s Settings
Ventilation
Settings
Alarm
Limits
Breath
Timeout
Source
Gas