CHAPTER 12: COMPLIANCE AND IEC CLASSIFICATION
IEC COMPLIANCE:
EN/IEC 60601-1-2 4th edition: 2014 (Medical electrical equipment Part 1-2: General
requirements for basic safety and essential requirements - Collateral standard: Electromagnetic disturbances
- Requirements and tests) defines the RF emissions limits and minimum RF immunity requirements for different
types of medical electrical equipment. The following tables include the RF emissions limits and RF immunity
requirements for, and results of tests upon, the Breathe Technologies® Life2000® Ventilation System.
The Breathe Technologies® Life2000® Ventilation System requires special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in
this
Instructions for Use
.
The use of accessories or cables other than those specified by Breathe Technologies® may result in increased
emissions or decreased immunity of the Breathe Technologies® Life2000® Ventilation System.
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC EMISSIONS
The Breathe Technologies® Life2000® Ventilation System is intended for use in the electromagnetic
environment specified below. The customer or user of the ventilation system should ensure that it is used in
such an environment.
Emission Test
Compliance
Electromagnetic Environment - Guidance
RF emissions CISPR 11
Group 1
The Breathe Technologies® Life2000® Ventilation System
RF emissions are within CISPR 11 guidelines and are not
likely to cause interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The Breathe Technologies® Life2000® Ventilation System
is suitable for use in all establishments, including domestic
establishments and those directly connected to the public
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC 61000-3-2
Complies
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies