19
2
STATIONARY CONFIGURATION
Turn the locking knob clockwise to the locked position to lock the
ventilator in the compressor.
When the compressor is on, check that the locked icon lights up
(green) to indicate the ventilator is properly docked. When the
ventilator is properly docked and locked the ventilator is being
charged by the compressor.
NOTES:
• If the compressor is off and connected to an AC power source,
press the battery charge status button on the compressor to
illuminate the green locked icon and verify that the ventilator is properly docked.
• If the knob does not lock on first attempt, turn knob slightly counterclockwise and then clockwise again.
• To ensure continuous ventilation, make sure the ventilator battery is being charged by the compressor
while the ventilator is docked. For more information see "Checking the Ventilator Battery Charge" on
page 23.
NOTE:
For undocking instructions, see "Undocking the Ventilator from the Compressor" on page 29.
VENTILATOR CHARGING BY COMPRESSOR (WHEN DOCKED AND LOCKED)
Compressor ON
Compressor OFF
AC CONNECTED to Compressor
Ventilator charging
Ventilator charging
AC DISCONNECTED from Compressor
Ventilator charging
Ventilator DOES NOT charge
NOTE:
The ventilator will not charge when the compressor is powered off and not connected to AC power.
0
100%
5
6
CONNECTING AN INTERFACE TO THE COMPRESSOR
Plug the Breathe Pillows Interface™ or Breathe Universal Circuit
™
connector
into the interface port on the side of the compressor until it clicks. For
more information about wearing interfaces, see "Chapter 5: Connecting
an Interface" on page 51.
0
100%
THE OUTLET FITTING IN STATIONARY CONFIGURATION
Do not connect a gas source to the outlet fitting on the compressor.
CAUTION:
Using the outlet fitting while the Life2000 Ventilation System is delivering therapy in Stationary
Configuration might decrease flow to the patient.
0
100%