158
|
Appendix B: Error Codes
Code
Description and Probable Causes:
4A0X
PMT gain out of range, x = readset.
Too much light in read chamber. Ensure the instrument enclosure is
completely installed and secured.
4B10
4B11
4B12
4B15
4B18
PMT measurement value is too low.
Failed high-voltage PMT test.
Failed low-voltage PMT test.
Failed well overload test for absorbance or fluorescence.
Failed background overload test.
Ensure that the Gen5 Filter Table accurately reflects the physical configuration
of the excitation and emissions filter wheels.
Ensure that the door is fully closed and there is not light leakage.
4E0X
4E11
4E12
Flash-on reference value at full scale during the flash fluorescence read
(x=readset #).
PMT test failed at 750 volts.
PMT test failed at 500 volts.
Too much light in chamber.
Door not closed completely.
Ensure that the Gen5 Filter Table accurately reflects the physical configuration
of the excitation or emission filter wheels.
Ensure that instrument case is completely installed and secured.
Sensitivity set too high; try adjusting sensitivity setting.
4E18
PMT saturation detected.
Too much light in chamber. Door not closed completely.
Ensure the Gen5 Filter Table accurately reflects the physical configuration of
the excitation and emission filter wheels.
Ensure the instrument case is completely installed and secured.
Sensitivity set too high. Try adjusting sensitivity setting.
Fluorescence standards dispensed to plate exceed value established by initial
standard values recorded.
Contamination within read chamber. Clean read chamber.
Verify there is no filter wavelength overlaps between excitation and emission
positions 2 and 3.
BioTek Instruments, Inc.
Summary of Contents for Synergy HTX
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Page 2: ......
Page 22: ...xviii Preface BioTek Instruments Inc ...
Page 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Page 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Page 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Page 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Page 166: ......
Page 174: ...150 Appendix A Specifications BioTek Instruments Inc ...