Fluorescence Specifications
|
147
Fluorescence Specifications
Read Timing
Because of the possible wide variations in setup, the following benchmark
conditions are specified:
Excitation Filter:
485/20 nm
Emission Filter:
528/20 nm
Measurements per data
point:
10
Delay after plate
movement:
100 ms
Minimum kinetic interval,
96-well
<55 seconds
Optical Probes
The Synergy HTX is configured with two probe sizes: The 3 mm probes can be
installed only in the top position, and the 5 mm probe can be installed only in the
bottom position.
Sensitivity
Specifications apply to the Normal mode of reading
5 mm optical probe
Bottom reading
DL Sodium Fluorescein in PBS
<= 30 pM
Excitation 485/20, Emission 528/20
DL Propidium Iodide in PBS
<= 62.5 ng/mL
Excitation 485/20, Emission 645/40
3 mm optical probe
Top reading
DL Methylumbelliferone in CBB
<=0.16 ng/mL
Excitation 360/40, Emission 460/40
Optional Time-Resolved Fluorescence (“T” models)
Delay:
0, or 20 to 2,000 µs
Integration Time:
20 to 2,000 µs
Granularity:
10-µs step
Synergy HTX Operator’s Manual
Summary of Contents for Synergy HTX
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Page 2: ......
Page 22: ...xviii Preface BioTek Instruments Inc ...
Page 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Page 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Page 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Page 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Page 166: ......
Page 174: ...150 Appendix A Specifications BioTek Instruments Inc ...