Error Codes
|
155
Code
Description and Probable Causes:
x=7: Plate width < 84.15 mm or > 86.11 mm
x=8: Plate length < 125.73 mm or > 128.40 mm
Review definition for plate defined in Plate Type database. Ensure counts or
dimensions do not exceed limits. See the Gen5 Help for a description of
measuring plates.
2B0x
Syringe failure.
x=1: Syringe failed to reach home sensor (optical sensor should be on).
x=2: Syringe moved off home sensor, but sensor didn’t changes state (optical
sensor should be off).
x=3: Syringe clean position too far from home sensor.
x=4: Steps to clear sensor at runtime deviated from value saved when homing
(verify error).
Protocol definition is incorrect or reader is being controlled by incorrectly
programmed third-party software.
Syringe was not installed correctly or was not cleaned, preventing a move to
home sensor.
Contact BioTek TAC.
2B0A
Plate not in carrier for system prime operation.
Place priming plate in carrier.
2C01
2C05
2C07
Dispenser configuration incorrect.
Volume calibration data is invalid.
Volume calibration is needed to override defaults.
Verify that the dispenser calibration values have been loaded into Gen5. See
the Installation chapter in the
Synergy HTX Operator’s Manual
.
Contact BioTek TAC.
2D09
Tip prime volumes specified could overflow the tip priming trough.
Empty the tip priming trough. Contact BioTek TAC.
2D0A
Tip prime trough or plate is full or may overflow.
Empty the tip priming plate.
2D15
Invalid kinetic interval selected for plate mode/plate synchronous mode.
Enter a valid kinetic interval. Call BioTek TAC.
2D16
Assay missed scheduled start of read (well synchronous mode).
Verify computer setup. Hibernate or sleep mode should not be enabled.
Contact BioTek TAC.
2D22
Invalid volume selected for tip prime.
Select a valid volume. Call BioTek TAC.
2D23
Invalid volume selected for dispense.
Select a valid volume. Call BioTek TAC.
Synergy HTX Operator’s Manual
Summary of Contents for Synergy HTX
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Page 2: ......
Page 22: ...xviii Preface BioTek Instruments Inc ...
Page 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Page 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Page 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Page 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Page 166: ......
Page 174: ...150 Appendix A Specifications BioTek Instruments Inc ...