8: Install the Dispense Module
|
19
white tip of the syringe plunger into the barrel of the syringe, and gently push
it all the way into the barrel.
11.
Install both syringes:
a.
Hold the syringe vertically with the threaded end at the top.
b.
Screw the top of the syringe into the bottom of the syringe valve. Finger-
tighten only.
c.
Carefully pull down the bottom of the syringe until it rests inside the hole
in the bracket.
d.
Pass a thumbscrew up through this hole, and thread it into the bottom of
the syringe. Hold the syringe to prevent it from rotating while tightening
the thumbscrew. Finger-tighten only.
12.
Locate the dispenser cable. Plug one end into the port on the left side of the
dispenser. Plug the other end into the dispenser port on the rear of the reader.
13.
Locate the injector-tip-cleaning stylus, packaged in a small cylinder. Attach the
cylinder to the back of the dispenser for storage.