Appendix B: Error Codes
Code
Description and Probable Causes:
5702
5703
Excitation filter wheel obstructed.
Emission filter wheel obstructed.
Ensure filter cartridge is fully inserted.
5706
Probe z-axis obstructed.
Shipping bracket not removed.
Verify plate height matches selected plate type.
Manually turn the z-axis up, remove any microplates from the carrier, and
attempt a successful system test.
5708
5709
Dispenser syringe 1 obstructed.
Dispenser syringe 2 obstructed.
Verify nothing is blocking the syringe drive.
5800
5801
Carrier x-axis obstructed.
Carrier y-axis obstructed.
Shipping screw still installed.
An object may be obstructing the carrier’s path.
5A00
5A01
Carrier x-axis obstructed.
Carrier y-axis obstructed.
Plate has hit something.
Plate cover not accounted for when creating plate dimension file.
Tip prime trough dislodged.
5B00
Plate height violation.
Plate is inside chamber when it should be outside.
•
The read was aborted and “home all axes” not performed.
•
The carrier is inside the reader and the newly defined plate height is
different from the most recently specified plate height. To resolve this
error, eject the carrier before running the experiment.
BioTek Instruments, Inc.
Summary of Contents for Synergy HTX
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Page 2: ......
Page 22: ...xviii Preface BioTek Instruments Inc ...
Page 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Page 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Page 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Page 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Page 166: ......
Page 174: ...150 Appendix A Specifications BioTek Instruments Inc ...