Chapter 6: Instrument Qualification
5.
The test report will appear; it will show either “SYSTEM TEST PASS” or
“SYSTEM TEST FAIL *** ERROR (error code) DETECTED.”
6.
Print the report, if required.
A sample test report is shown on the next few pages.
Gen5 stores the results in a database, so the results can be
retrieved/printed at any time. We recommend that you print and save
the reports to document that the test was performed.
7.
If the test failed, look up the error code in
Appendix B, Error Codes
to
determine its cause. If the cause is something you can fix, turn off the
reader, fix the problem, and then turn the reader back on and retry the test.
If the test continues to fail, or if the cause is not something you can fix,
contact BioTek’s Technical Assistance Center (see page 6 for contact
information.)
8.
Turn off the incubator.