15: Test the Injector System
|
25
The dispense module’s setup should resemble the photo in Figure 11.
Make any final adjustments, if necessary.
Figure 11: The fully assembled dispense module
5.
Select
System > Instrument Control > Synergy HTX
.
6.
Click the
Prime
tab.
7.
With
Dispenser
set to
1
, set the
Volume
to
5000 µL
and click
Prime
.
The syringe should move down and up repeatedly, drawing fluid from the bottle.
The fluid should pump through the tubing and dispense into the priming plate.
Examine the fittings; no leaks should be detected.
If leaks are detected, tighten all fittings and repeat the prime. If leaks are still
detected, contact BioTek’s Technical Assistance Center.
8.
When the prime finishes, set
Volume
to
2000 µL
and click
Purge
to clear the fluid
lines.
9.
Set
Dispenser
to
2
and repeat steps 7 and 8.
10.
When finished, remove and empty the priming plate.
11.
Close the software.
Synergy HTX Operator’s Manual