130
|
Chapter 6: Instrument Qualification
1
2
3
4
5
6
7
8
9
10
11
12
A
B
C
100 nM
50 nM
25 nM
12.5 nM
6.25 nM
BUF
D
100 nM
50 nM
25 nM
12.5 nM
6.25 nM
MUB
100 nM
BUF
E
100 nM
50 nM
25 nM
12.5 nM
6.25 nM
BUF
F
100 nM
50 nM
25 nM
12.5 nM
6.25 nM
G
H
Protocol Reading Parameters
Your tests may require modifications to some of the parameters below,
such as Plate Type or Gain (see “Troubleshooting” on page 120). The
Plate Type setting should match the plate you are actually using.
The following table contains the recommended settings for the
Synergy HTX
FI_MUB.prt
protocol.
Protocol Name: Synergy HTX FI_ MUB.prt
Parameter
Default Setting
Plate Type:
Costar 96 black opaque (Mfr. #3915)
Read Step 1
Kinetic Loop:
Run time :45 seconds
Interval: 3 seconds
Reads: 16
Read Step Label:
“Sensitivity Read”
Read Well:
D7
Filter Set:
EX 360/40 nm
EM 460/40 nm