Chapter 4: Preventive Maintenance
Removing the Reader’s Shroud
Caution! Before removing the shroud:
Purge the
dispense module (see page 57 for instructions), and then
turn off and disconnect the reader from its power supply,
the PC, and the dispense module.
The reader’s shroud (cover) must be removed to expose the internal components.
1.
If you have not already done so, purge the dispense module of fluid.
2.
Clear the work surface around the reader so you can easily access all sides
of the instrument.
3.
Disconnect power and all cables. Set the dispense module aside.
4.
Remove four black mounting screws: one at the bottom rear corner on each
side, and two at the top center of the rear panel.
When reinstalling the shroud, press down firmly on the top to maintain
a good seal while tightening the top screws.
Remove two
side screws
Remove two screws from the
top center of the rear panel