Routine Procedure for Models with Injectors
|
87
Rinse the Fluid Lines
Perform this procedure only if decontamination was performed using sodium hypochlorite.
1.
Place a beaker containing at least 30 mL of deionized water on the dispense
module.
2.
Place the SYRINGE 1 or 2 inlet
tube in the beaker.
3.
If you have not already done so, place the outlet
tubes in an empty beaker.
4.
From the Reader
Control dialog, select
Dispenser 1
or
2
, set the Volume to
5000 µL
, and keep the default dispense
Rate
.
5.
Run five prime cycles, for a total of 25000
µ
L.
6.
Pause for 10
minutes and then run one prime cycle with 5000 µL. This delay
will allow any residual sodium hypochlorite to diffuse into the solution and
be flushed out with the next prime.
7.
Empty the beaker containing the outlet tubes.
8.
Wipe all surfaces with deionized water.
9.
Discard the used gloves and cloths using a biohazard trash bag and an
approved biohazard container.
Clean the Internal Tubing and Injector Heads
Turn to
Chapter 4, Preventive Maintenance
and perform the following
procedures to access, remove, and clean the internal tubing and injector heads:
•
Required Materials
•
Removing the Reader’s Shroud
•
Removing the Internal Tubes and Injector Heads
•
Cleaning the Internal Tubes and Injector Heads
When finished, replace the internal components and the reader’s shroud.
Synergy HTX Operator’s Manual
Summary of Contents for Synergy HTX
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Page 2: ......
Page 22: ...xviii Preface BioTek Instruments Inc ...
Page 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Page 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Page 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Page 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Page 166: ......
Page 174: ...150 Appendix A Specifications BioTek Instruments Inc ...