Models with Injectors
|
149
Models with Injectors
Dispense/Read
Specifications apply to models with the dual-reagent dispense module
Plate Type:
Dispenses to standard 6-, 12-, 24-, 48-, 96-, and 384-well
microplates with 128 x 86 mm geometry
Detection Method:
Absorbance, Fluorescence, Luminescence, TRF
Volume Range:
5-1000 µL with a 5-20 µL tip prime
Accuracy:
± 1 µl or 2.0%, whichever is greater
Precision:
Dispensing a 200 µL solution of deionized water,
0.1% Tween 20, and dye at room temperature:
< 2.0% for volumes of 50-200 mL
< 4.0% for volumes of 25-49 mL
< 7.0% for volumes of 10-24 mL
< 10.0% for volumes of 5-9 mL
Injection Speeds:
225, 250, 275, and 300 µL/sec
Maximum Delay
between End of
Dispense Process
and Start of Read
Process (96-/384-
well plates, default
probe heights only)
Absorbance
: T ≤
3 sec
Top Filter Fluorescence
: T ≤ 1 sec
Bottom Filter Fluorescence
: T ≤ 1 sec
Luminescence
: T ≤ 0.5 sec
Synergy HTX Operator’s Manual