Dispense Module Tests
|
139
3.
Create
Data Reduction
steps to calculate Delta OD values:
•
Select
Protocol > Data Reduction
and select
Custom
.
•
Within this dialog, click the
Select Multiple Data Sets
button and then
click the
DS2
radio button.
Set the
Data In
for
DS1
to the
80
µL Read step at
405
nm.
Set the
Data In
for
DS2
to the
80
µL Read step at
750
nm.
•
Click
OK
to return to the dialog.
•
In the
New Data Set Name
field, type an identifying name such as ‘Delta
OD 80 uL_Disp 1’.
•
Clear
Use single formula for all wells.
•
In the
Current Formula
field, type
DS1-DS2
and then assign the formula
to wells
A1
to
H4
.
•
Click
OK
to add the transformation to the Data Reduction list.
•
Create another Transformation similar to the above, with these
characteristics:
DS1
set to the
20 and 5
µL Read step at
630
nm
DS2
set to the
20 and 5
µL Read step at
750
nm
New Data Set Name
resembling ‘Delta OD 20 and 5 uL_Disp <#>’
Formula
DS1-DS2
applied to wells
A5
to
H12
4.
(This step is optional.) The results analysis worksheet at the end of this chapter
requires the calculation of the Standard Deviation, Mean, and % CV of the ODs
read for each dispense volume in each plate (six sets of calculations). By
identifying the wells by their dispense volumes in the Plate Layout, Gen5 will
calculate these values for you.
•
Open the Plate Layout dialog.
•
Define three Assay Control IDs and assign them to the following wells:
Disp_80
A1 to H4
Disp_20
A5 to H8
Disp_5
A9 to H12
After running the experiment, view the Statistics for each Delta OD
Data Set to view the calculations.
5.
Save the protocols as
Synergy HTX
Disp 1 Test.prt
and
Synergy HTX
Disp
2 Test.prt
.
Gen5 Test Protocols for Models without Absorbance Capabilities
The test procedure on page 135 dispenses three volumes of fluid to a microplate and then
reads the plate on an absorbance reader. The procedure is performed twice, once for each
Synergy HTX Operator's Manual
Summary of Contents for Synergy HTX
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Page 2: ......
Page 22: ...xviii Preface BioTek Instruments Inc ...
Page 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Page 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Page 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Page 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Page 166: ......