94
|
Chapter 6: Instrument Qualification
Recommended Qualification Schedule
The schedule below defines the factory-recommended intervals for qualifying a Synergy
HTX used two to five days a week. The actual frequency, however, may be adjusted
depending on your usage of the instrument. This schedule assumes the reader is properly
maintained as outlined in
Chapter 4, Preventive Maintenance
.
Tests
IQ
OQ
PQ
Initially
Initially/
Annually
Monthly
Quarterly
System Test
Absorbance Tests
*
Absorbance Plate Test
Liquid Test 1 or Liquid Test 2
**
Liquid Test 3
***
Fluorescence Tests
Corners Test
Sensitivity/Linearity Tests
Tests for Injector Models
Injector System Test
Dispense Accuracy and
Precision Tests
*
If applicable to your reader model
**
Regarding Liquid Tests 1 and 2:
•
If you have an Absorbance Test Plate, run Liquid Test 1.
•
If you do not have an Absorbance Test Plate, run Liquid Test 2.
***
Liquid Test 3 is optional; it is provided for sites requiring verification at
wavelengths lower than those attainable with the Absorbance Test Plate.
Important!
The risk factors associated with your assays may
require that the Operational and Performance Qualification
procedures be performed more frequently than shown above.
BioTek Instruments, Inc.
Summary of Contents for Synergy HTX
Page 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Page 2: ......
Page 22: ...xviii Preface BioTek Instruments Inc ...
Page 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Page 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Page 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Page 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Page 166: ......
Page 174: ...150 Appendix A Specifications BioTek Instruments Inc ...