BioTek Instruments, Inc.
Number of Standard Replicates
This option allows you to enter the number of replicates per standard group in
the assay. Any predefined value appears on the display.
E N T E R N U M B E R O F
S T A N D A R D R E P L I C A T E S :
0 2
•
Use the
numeric
keys to enter the
NUMBER OF STANDARD
REPLICATES
. The range is 1 to 8 replicates. The software will verify
that the number of replicates, multiplied by the number of standards,
does not exceed the number of wells on the plate.
•
Press
CLEAR
to clear the value on the display.
•
Press
ENTER
to continue.
Average Standards
The
AVERAGE STANDARDS
option allows you to select whether or not to
average the replicates of each standard group. This average is used to calculate
the standard curve instead of using the individual replicates of each standard.
If the number of standard replicates is 1, this option is not available.
A V E R A G E S T A N D A R D S ? Y E S
Y E S N O
•
Select
YES
to average the replicates for each standard group, and then
use the group averages when calculating the standard curve.
•
Select
NO
to use the individual standard replicates when calculating the
standard curve.
•
Press
ENTER
to continue.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...