Qualification Procedures
|
107
ELx800 Operator's Manual
Checksum Test
This test also runs automatically when the reader is turned on. The test compares
the software to the internally recorded checksum values to ensure that the
programming has not been corrupted. It verifies the checksum, and displays the
part numbers and versions of software currently loaded onto your reader. If there
are any errors during the power-on checksum test, they will be displayed.
1.
Press
UTIL
at the Main Menu. At the
SELECT UTILITY OPTION
screen,
press
TESTS
and then
CHKSUM
at the
SELECT TEST
screen to run the
Checksum Test. The information displayed will
resemble
the following:
7 3 3 0 2 0 2
V E R S I O N 3 . 0 7
C O D E C H E C K S U M :
( 1 7 A 5 )
The initial checksum test display will show the
onboard (base code)
software
part number, version number, and checksum. After a few
moments, a second screen will display:
7 3 3 0 2 9 1 - F W
V 1 . 0 0 . 0
The second checksum test display will show the
assay configuration
software part number and version number. After a few moments, the
Main Menu will reappear.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...