116
|
Chapter 4: Instrument Qualification
BioTek Instruments, Inc.
Solution B
Required Materials:
•
BioTek QC Check Solution No. 1 (PN 7120779, 25 ml; or 7120782,
125 ml)
•
Deionized water
•
5-ml Class A volumetric pipette
•
100-ml volumetric flask
Preparation of Stock Solution:
1
Pipette a 5-ml aliquot of BioTek QC Check Solution No. 1 into a 100-ml
volumetric flask.
2
Make up to 100 ml with DI water; cap and shake well.
This should create a solution with an absorbance of about 2.000 when
using 200 µl in a flat-bottom microwell. The OD value result will be
proportional to the volume in the well and the amount of QC Check
Solution No. 1 used. You can use a larger or smaller well volume, or
add more Check Solution or water to adjust the stock solution. Note
that too small a well volume may result in increased pipetting-related
errors.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...