Filter Storage and Handling
|
129
ELx800 Operator's Manual
Tools and Supplies
•
Mild detergent
•
Deionized or distilled water
•
Clean, lint-free cotton cloths
Procedure
1.
Turn off and disconnect the instrument from the power supply.
2.
Moisten a clean cotton cloth with water, or with water and the mild detergent.
Do not soak the cloth.
3.
Wipe the plate carrier and all exposed surfaces of the instrument.
4.
If detergent was used, wipe all surfaces with a cloth moistened with water.
5.
Use a clean, dry cloth to dry all wet surfaces.
Filter Storage and Handling
To properly store interference filters for the ELx800™ microplate reader during extended
periods of non-use, package the filters in a light-tight envelope or container, away from
high humidity. This will ensure the longest life for the filters. When handling the filters,
keep the surfaces clean from fingerprints and debris by simply wiping with a lens tissue or
other lint free cloth.
Summary of Contents for EL800
Page 1: ...Operator s Manual Absorbance Microplate Reader ELx800 ...
Page 2: ... ...
Page 4: ... ...
Page 62: ...36 Chapter 2 Installation BioTek Instruments Inc ...
Page 126: ...100 Chapter 3 Operation BioTek Instruments Inc ...
Page 162: ...136 Chapter 5 Maintenance and Decontamination BioTek Instruments Inc ...
Page 182: ...156 Appendix A Computer Control BioTek Instruments Inc ...
Page 194: ...168 Appendix B Using 384 Well Geometry BioTek Instruments Inc ...
Page 197: ... 171 ELx800 Operator s Manual Figure 39 Curve Fit Report ...
Page 199: ... 173 ELx800 Operator s Manual Figure 41 Column Report without samples ...
Page 200: ...174 Appendix C Report Format BioTek Instruments Inc Figure 42 Panel Report ...
Page 201: ... 175 ELx800 Operator s Manual Figure 43 Assay Detail Report Sheet 1 of 2 ...
Page 204: ...178 Appendix D Comparison of the ELx800 and the EL800 BioTek Instruments Inc ...
Page 218: ...192 Appendix E Instructions for Programming a New Assay BioTek Instruments Inc ...